Saturday, September 17, 2011

Java Applet of Radiography Principles


1) Java Simulation 
(Please remember to allow running of scripts/ActiveX controls for viewing the applet)


  • INTENSITY-DISTANCE INVERSE SQUARE LAW


The intensity-distance inverse square law is given by




where I1 and I2 are the intensities at 2 points and  d1 and dare their respective distances from the source. The law assumes a fixed source.


  • SFD-EXPOSURE FORMULA


The sfd-exposure formula is given by




where E1 and E2 are the respective exposures required to produce the same quality of image when the film is at distances SFD1 and SFD2 from the source.  An exposure can be changed in 3 ways:


1) Varying the source intensity, Is, keeping the exposure time, t, fixed.
2) Varying the exposure time, t, keeping the source intensity, Is, fixed.
3) Varying both the source intensity, Is, and the exposure time, t.
                             
In the Java simulation, a constant source intensity(Is1 = Is2) is assumed and since
The respective exposure times, t1 and t2 , required at the source-to-film distances, SFD1 and SFD2,  are given by the formula,


                                    


2) Theory
                                

  • A Conventional X-Ray Unit

Figure 1  Conventional design of glass envelope X-ray tube

X-rays are a form of electromagnetic radiation, of the same physical nature as visible light, radiowaves, etc. However, they have a wavelength which allows them to penetrate most materials with partial absorption during transmission. Their wavelength varies from about 10 nm for low energy radiation to about 10-4 nm for high energy x-rays which will penetrate up to 500 mm in steel.

X-rays are generated when a beam of high energy electrons is stopped suddenly by a metal target. The essentials of an X-ray tube are shown in Figure 1.

To produce x-rays, the filament is heated with a current until it is hot enough to emit electrons. The tungsten target or anode is charged positively with respect to the filament and therefore the negatively charged electrons are attracted to the anode. At the anode, the electrons are stopped and they then produce x-rays which are directed towards the object to be radiographed.

Three variables which can be adjusted using the instrument panel connected to the x-ray tube are energy (in kV), intensity (in mA) and the exposure time (in min) which the x-rays are left on for during an x-ray shot. These variables need to be adjusted accurately in order to produce a satisfactory radiograph.

The high voltage difference between the filament and the target is termed the kV used for the radiograph. Typical values in industrial radiography are 50 to 300 kV although higher and lower values are used. If the kV is higher, the x-rays produced have more energy and  therefore they can penetrate a thicker component. The general practice is to increase the kV used as the thickness of the component to be radiographed increases. The kV can be considered as the quality of the x-rays.

The mA is a measure of the electron flow striking the target. The focusing cup can control the mA and also focus the electrons onto the target. The greater the mA value, the greater the quantity of x-rays produced at the target. Therefore the mA is a measure of the amount of x-rays leaving the target. This is also called the intensity.

The Exposure (in mA.min), E, given to a film is equal to the product of the intensity, I, and the exposure time, t, i.e., E = I x t.

At the anode, only a small proportion (1-10%) of the energy of the electrons is converted to x-rays and most becomes heat energy. The tungsten target therefore needs to be air, water or oil cooled.

The effective width of the source of the x-rays is considerably smaller than the area of the target on which the electrons are incident. Effective source widths vary up to 5 mm in diameter.


  •  Production of A Radiograph


Figure  2  Arrangement for film radiography

A radiograph is a photographic image produced by a beam of penetrating ionised radiation after passing through a specimen and radiography is the production of radiographs.

The usual arrangement for producing a radiograph is shown in Figure 2, using a small diameter source G, and a sheet of film as a detector.

The cavity in the specimen, as shown at B, causes a lower absorption along the path GBF. More radiation reaches the film at point F, compared with, say, point  N. Therefore an x-ray "image" of the cavity is produced.

To produce a radiograph, the x-rays are allowed to reach the film for an appropriate exposure time, which depends on the intensity of the x-rays, the thickness of the specimen, and the characteristics of the film. The film is then processed (developed, fixed, washed and dried) and the defects can be seen as blackened areas. The film is then placed on an illuminated screen so that the image can be examined and interpreted.  



  •  Intensity-Distance Law 


Figure 3  Intensity varies inversely to the distance from source

As ionised radiation travels in straight lines outward from a source, i.e., it is not focused, the simple "inverse square law" applies since when the distance between the film and the source is increased, the radiation has to cover a larger area and so is reduced in intensity as shown in Figure 3.  Assuming a fixed source,  intensities I1 and I2 at two points which are at distances d1 and d2 from the source, are related as,


Therefore,

A Java simulation of the inverse square law can be found at this link> Intensity-Distance Law.   .  



  •  SFD-Exposure Formula  

If a film is exposed at a distance "D" from the x-ray tube and a certain density produced, then to produce the same density on a similar film at a distance of "2D", a greater exposure is required. The exposure required to produce the same quality of image is related to the source-to-film distance(sfd) by, 



where E1 and E2 are the respective exposures required when the film is at distances SFD1 and SFD2 from the source.  

An exposure can be changed in 3 ways:
1) Varying the source intensity, Is, keeping the exposure time, t, fixed.
2) Varying the exposure time, t, keeping the source intensity, Is, fixed.
3) Varying both the source intensity, Is, and the exposure time, t.


In the Java simulation, a constant source intensity(Is1 = Is2) is assumed and since         

The respective exposure times, t1 and t2 , required at the source-to-film distances, SFD1 and SFD2,  are given by the formula,


Therefore,

The Java simulation of the sfd-exposure formula can be found at this link> SFD-Exposure Formula.  

No comments:

Post a Comment